PR233 Series Portable Process Calibrator

FEATURES:

  • The analog resistor/thermal resistor function is highly adaptable and can be a circular chart recorder.
  • A secondary instrument such as a balance recorder provides a stable signal.
  • The measurement and output functions are isolated from each other and can be used simultaneously.
  • Industrial-grade resistive touch screen + mechanical button human-computer interaction operation, can Intuitive and convenient setup via touch screen, or mechanical press The keys perform common operations.
  • The resistance measurement function can select three-wire, three-wire and four-wire resistance measurement methods.
  • Built-in high-precision reference temperature compensation and thermocouple quick interface (PR23C does not have this function)
  • Recording function, can realize on-site paperless recording
  • Built-in isolated 24V power supply for easy verification of temperature transmitters

  • Share on

PR233 Series Portable Process Calibrator

PR233 series portable process calibrator has simultaneous measurement and analog voltage, resistance, current, thermocouple, thermal resistance, frequency, switching and other signals

The function. Compared with the PR231 series multi-function calibrator, the PR233 series products are very simple in function and operation. They are compact in design and easy to carry, which is very suitable for field applications.

Basic Parameters

Weight

510g

Charging   power source

100~240V AC,   50~6HZ

Size

175mm×90mm×52mm

Working Temp.

-10℃~50℃

Battery type

7.8V 2200mAh   Rechargeable lithium battery

Charging   temp.

5℃~35℃

Working   hours

10~15h, Loop   power off

Humidity

0~80%,Non   condensing

4~6h, Loop   power on

Preheating   time

Preheat 10   minutes

Charging   time

3 hours

Calibration   period

1 year

What is process calibration

Process industries are those in which the final product, once manufactured, can’t be separated into its individual components. Examples include the pharmaceutical, petrochemical, electrical power, and semi-conductor industries. Calibrations performed within these industries may happen on a bench in a calibration laboratory; they can also happen out in the plant or in the field. We refer to both bench and field calibration practices that happen within process industries as “process calibration.”

Application

Process calibration professionals can include metrologists, instrumentation and control (I&C) or electrical and instrumentation technicians, electrical technicians, or engineers. Some of these professionals work in a metrology lab. Others work in an instrument shop, and still others work out on the plant floor.

 

Source Accuracy

Function

Range

Measurement   range

Resolution

Accuracy

Remarks

Voltage

100mV

-5mV~120mV

0.1uV

0.005%RD+5uV

Input   impedance

≥50mΩ

1V

-50mV~1.2V

1uV

0.005%RD+0.005%FS

50V

-0.5V~50V

0.1mV

Input   impedance

≥1mΩ

Current

50mA

-5mA~50 mA

0.1uA

0.005%RD+0.005%FS

Internal   resistance =10Ω

Resistance

500Ω

0Ω~500Ω

1mΩ

0.005%RD+0.005%FS

Output   1mA current

5kΩ

0kΩ~5kΩ

10mΩ

Output 0.1mA   current

Thermocouple

S,R,B,K,N,J,E,T,EA2,Wre3-25,Wre5-26

0.1℃

Note

According   to ITS-90 scale

Reference

internal

-10℃~60℃

0.01℃

0.3℃

Thermal   resistance

PT10,   Pt100,Pt200,Cu50,Cu100, BA1, BA2, JPt100, Pt500, Pt1000

0.01℃

Note

ρ-value

50S

0.001%~99.999%

0.001%

0.005%

 

Input   pulse width amplitude range:1V~50V

Frequency

1kHz

0.00001kHz~1.2 kHz

0.01Hz

0.01%FS

100kHz

0.01kHz~120 kHz

10Hz

0.1%FS

Main   range temperature characteristics

100mV

1V

50V

50mA

500Ω

5kΩ

Need to   enter the correction value

0.3uV/℃

3uV/℃

10uV/℃

0.5uV/℃

0.7mΩ/℃

7mΩ/℃

Measurement Accuracy

Function

Range

Measurement range

Resolution

Accuracy

Remarks

Voltage

100mV

-5mV~120mV

0.1uV

0.015%RD+10uV

Max load curren=7mA

1V

-0.2V~1.2V

10uV

0.015%RD+0.005%FS

10V

-2V~12V

0.1mV

Current

20mA

0mA~25mA

1uA

0.005%RD+0.005%FS

Max load voltage=24V

Resistance

350Ω

0Ω~350Ω

1mΩ

Note

According to ITS-90 scale

4kΩ

0kΩ~4kΩ

10mΩ

Thermocouple

S,R,B,K,N,J,E,T,EA2,Wre3-25,Wre5-26

0.1℃

Note

Thermal resistance

PT10, Pt100,Pt200,Cu50,Cu100, BA1, BA2, JPt100,   Pt500, Pt1000

0.01℃

Note

Frequency

1kHz

0.00001kHz~1.2 kHz

0.01Hz

0.01%FS

Max load curren=7mA

100kHz

0.01kHz~120 kHz

10Hz

0.1%FS

24V output

Max voltage error:0.3V Ripple noise:35mVp-p(20MHZ   bandwidth)

Max load curren: 70mA Load regulation: 0.5%(load   10%~100% change)

 

Main range temperature characteristics

100mV

1V

50V

50mA

500Ω

5kΩ

Need to enter the correction value

0.5uV/℃

5uV/℃

50uV/℃

1.5uV/℃

1.5mΩ/℃

15mΩ/℃