General Specifications (Unless otherwise indicated all specifications are rated from a nominal 23 °C, 70 % RH for 1 year from calibration) |
|
Operating Temperature Range | -20 to 60 °C (-5 to 140 °F) |
Storage Temperature Range | -30 to 60 °C (-22 to 140 °F) |
Relative Humidity Range | 10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing |
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing | |
Size | 7.00 X 3.30 X 2.21 inches (177.8 x 83.8 x 56.1mm) |
Weight | 14.0 oz (397 grams) |
Battery | 4 AA Alkaline, 120 VAC 50/60 Hz AC adaptor included |
Miscellaneous | Low battery indication with nominal 1 hour of operation left |
Over-voltage protection to 120 Vrms (rated for 30 seconds) or 240 Vrms (rated for 15 seconds) | |
Bar graph display with 1% resolution of 4-20/10-50 mA signal scale | |
High contrast graphic liquid crystal display with 0.45” (11.4 mm) high digits | |
Model 535A Analog Output | 10 mV/mA ±(0.1% + 0.04mV), 5 ppm/°C |
Common Specifications for all Current Modes: | |
Ranges | 0.000 to 24.000 mA, -25.00 to 125.00% of 4-20 mA |
0.000 to 52.000 mA, -25.00 to 105.00% of 10-50 mA | |
Accuracy | ≤ ± (0.025 % of reading + 0.004 mA) |
Temperature Effect | ≤ ± 50 ppm/°C of range |
Resolution | 0.001 mA and 0.01 % |
Step/Ramp Timebase Accuracy | 0.01% of 4.9152 MHz |
Source/Power and Measure 2-Wire Transmitter Specifications: | |
Loop Compliance Voltage | ≥ 43 Volts |
Loop Drive Capability | 1200 Ω at 20 mA/800 Ω at 50 mA for entire battery life |
Miscellaneous | Open loop or out of compliance conditions are indicated by appropriate error display |
Battery life in: | |
Source mode ≥ 18 hrs at 12 mA/≥ 9 hrs at 30 mA nominal (100 W load) | |
Power measure ≥ 10 hrs at 12 mA/≥ 5 hrs at 30 mA nominal (100 W load) | |
Selectable EZ-Step(s) for Source Mode/2-Wire Transmitter Simulation: | |
2 to 16 selectable step settings | |
Step size is determined by the selected high & low ranges | |
Selectable time settings for stepping and soak: | |
STEP: 5 to 900 seconds SOAK: 0 to 900 seconds | |
Read mA Specifications: | |
Voltage Burden | ≤ 2V |
Overload/Current Limit Protection | ≤ 24 mA (4-20 range)/ ≤ 52.5 mA (10-50 range) nominal |
Battery Life | ≥ 40 hours typical |
2-Wire Transmitter Simulation Specifications: | |
Overload/Current Limit Protection | ≤ 24 mA (4-20 range)/ ≤ 52.5 mA (10-50 range) nominal |
Loop Voltage Limits | 2-90 VDC |
Miscellaneous | Open loop or out of compliance conditions are indicated by appropriate error display |
Battery life ≥ 40 hour typical | |
Selectable EZ-Step(s) for Source Mode/2-Wire Transmitter Simulation: | |
2 to 16 selectable step settings | |
Step size is determined by the selected high & low ranges | |
Selectable time settings for stepping and soak: | |
STEP: 5 to 900 seconds SOAK: 0 to 900 seconds | |
Read Voltage Specifications: | |
Range: | 0.00 to 30.00 VDC (with 4X over range) |
Resolutions | 0.01 VDC |
Temperature Effect | ≤ ± 200 ppm/°C of range |
Input Resistance | ≥ 1 MΩ |
Accuracy | 0.00 to 30.00 VDC ≤ ± (0.1 % of reading ±0.1 V) |
Source Voltage Specifications: | |
Source Range: | 0.000 to 24.000 VDC |
Output Resistance | ≤0.3 Ω |
Source Current | ≥20.000 mA |
Accuracy | ≤±(0.025%RDG + 0.004 V) ±50ppm/°C of range |
Ordering Information: | |
4-20/10-50 Milliamp Dual Range Loop Calibrator | Model 535 |
Includes: Calibration Test Data NIST Traceable Certificate Carrying Case 60 Hz AC adapter 50 Hz AC adapter |
020-0200 (included) 020-0100 (included) 020-0101 (available upon request) |
Option: 10 mV/mA analog output | Model 535A |