The 8100 Series Multimeter signals a revolution in the way that precision measurements are performed

  • 4 / 9 PPM Accuracies
  • Up to 30A Current Ranges
  • Resistance Measurement to 1 TOhm
  • Advanced Maths Functions
  • Full Colour Screen
  • USB / RS232 / GPIB / LAN Interfaces
  • Dedicated Low Noise Electrometer Input (8104)
  • Rear Panel Inputs (8104)

  • Share on

INTRODUCTION

The 8100 Series from Transmille is the most advanced multimeter available on the market. Featuring a full colour screen, the 8100 Series offers a wide range of functionality and covers a wide range of input parameters whilst achieving breakthrough performance. The 8100 is a significant improvement on previous technologies and remains at an affordable price point.

HIGH PERFORMANCE DESIGN

The 8100 Series is aimed at laboratories, manufacturers and design facilities requiring a wide range of measurement parameters with specifications as low as 4ppm. The 8100 is designed with metrologists in mind, with great care taken in the design of sensitive input electronics.

Ultra low input bias current on DC Voltage inputs minimise errors when utilised as a null meter with Resistive Dividers, and High Input Impedance on AC Voltage up to 10V reduces errors when measuring current through use of current shunts to enable the 8100 Series ensuring reliable measurements across multiple functions.
Dynamic display of accuracy, 8.5 digital precision,  measurement uncertainty, standard deviation and more provide metrologists with immediate feedback of their measurement and enable users to maximise performance while improving confidence in measurements. 

ADVANCED MEASUREMENT FUNCTIONS

The 8100 Series has been optimised for use with Transmille’s range of AC/ DC current shunts enabling DC and offset parameters to be entered in providing direct measurement of applied current on the screen. The 8100 simultaneously measures the voltage and frequency and performs a real time correction based upon the AC/DC difference at the current frequency and the value of the shunt.

The high input impedance of the 8100 series below 10V AC removes the requirement for buffer amplifiers minimising sources of errors and simplifying connections.

 

8104

8109

DC VOLTAGE 0 - ±1025V
±4ppm
0 - ±1025V
±9ppm
DC CURRENT 0 - ±30A
±7ppm
0 - ±30A
±25ppm
DC CURRENT (ELECTROMETER INPUT) 10nA - ±10uA Ranges, 0.01pA Resolution
±30ppm
N/A
AC VOLTAGE 20mV - 1025V, 10Hz - 1MHz
±0.015%
20mV - 1025V, 10Hz - 100kHz
±0.05%
AC CURRENT 0 - 30A
±0.03%
0 - 30A
±0.08%
RESISTANCE 0 - 1 GOhms
±8ppm
0 - 1 GOhms
±20ppm
RESISTANCE (ELECTROMETER) 10 MOhms - 1T Ohm, 10V - 300V
±140ppm
N/A
FREQUENCY 10Hz - 1MHz
±5ppm
10Hz - 1MHz
±5ppm
TEMPERATURE (PRT / RTD) 2, 3 & 4 Wire
Support for IEC 751 & ITS 90 Coefficients
N/A
TEMPERATURE (THERMOCOUPLE) Support for over 8 Types
±0.05°C
N/A
REAR PANEL TERMINALS Fitted as standard N/A