• The most complete IEC 60601 tester on the market;
• 10 universal configurable connections. They can be configured as applied parts, non-earthed parts, earthed parts;
• Dedicated test ports for connection to functional earthing and signal I/O connections;
• All possible configurations acc. to the IEC 60601 and IEC 62353 standards;
• Tests can be run on the instrument directly or via a PC;
• Fully automatic test flow with Auto Sequences®;
• Tests and limits are automatically set acc. to the set configuration of medical equipment;
• Test sequences are optimized for fastest work flow;
• Supports measuring by standards IEC/EN 60601 (2nd and 3rd editions), IEC/EN 62353, ANS/NZS 3551, EN 50678 and EN 50669;
Category: Medical Supply
Tags: Metrel Indonesia, Medical Tester, Agen Metrel Indonesia, Terminal Elektronika Sekawan
The MI 6601 MediTest can be used as stand-alone tester without a PC or laptop in the field. For office and laboratory use, the tester can work in combination with Metrel Medical Electrical Safety Manager (MMESM). It supports creation and execution of test sequences in accordance with IEC/EN 60601 and asset management. Reports are created with integrated online services Metrel Cloud Reports and Metrel Cloud Storage. Compliance with IEC/EN 60601 is a widely accepted benchmark and a requirement for commercialisation of electrical medical equipment around the world. In Metrel we believe that our new MI 6601 MediTest is the fastest and easiest-to-use electrical safety compliance tester on the market, which entirely covers prescribed standard test procedures, without any compromises or simplifications.
TECHNICAL SPECIFICATION |
||||
FUNCTION | MEASURING RANGE | RESOLUTION | ACCURACY | |
Continuity / Protective earth resistance ^{1} Continuity |
R | 0.00 Ω … 19.99 Ω | 0.01 Ω | ±(2 % of reading + 2 D) |
20.0 Ω … 99.9 Ω | 0.1 Ω | ±3 % of reading | ||
100.0 Ω … 199.9 Ω | 0.1 Ω | ±5 % of reading | ||
200 Ω … 999 Ω | 1 Ω | Indicative | ||
Insulation Resistance (Riso, Riso-S) ^{2} Insulation resistance, Insulation resistance –S (250 V, 500 V) |
Riso | 0.00 MΩ … 19.99 MΩ | 0.01 Ω | ±(3 % of reading + 2 D) |
20.0 MΩ … 99.9 MΩ | 0.1 Ω | ±5 % of reading | ||
Riso-S | 100.0 MΩ … 199.9 MΩ | 0.1 Ω | ±10 % of reading | |
^{2} Output voltage | Um | 0 V … 600 V | 1 V | ±(3 % of reading + 2 D) |
Sub-Leakage Current, Substitute Leakage Current - S ^{3} Substitute leakage current, Substitute leakage current - S |
Isub | 0.00 mA … 1.99 mA | 0.01 mA | ±(3 % of reading + 3 D) |
Isub-S | 2.00 mA … 19.99 mA | 0.01 mA | ±5 % of reading | |
Differential Leakage current ^{4} Differential leakage current |
Idiff | 0.000 mA … 1.999 mA | 1 µA | ±(3 % of reading + 3 D) |
2.00 mA … 19.99 mA | 0.01 mA | ±5 % of reading | ||
^{4} Power (active) | P | 0 W … 999 W 1.00 kW … 3.70 kW |
1 W 10 W |
±(5 % of reading + 5 D) ±5 % of reading |
PE leakage current ^{5} PE leakage current |
Ipe | 0.000 mA … 1.999 mA | 1 µA | ±(3 % of reading + 3 D) |
2.00 mA … 19.99 mA | 0.01 mA | ±5 % of reading | ||
^{5} Power (active) | P | 0 W … 999 W 1.00 kW … 3.70 kW |
1 W 10 W |
±(5 % of reading + 5 D) ±5 % of reading |
Touch leakage current ^{6} Touch leakage current |
Ipe | 0.000 mA … 1.999 mA | 1 µA | ±(3 % of reading + 3 D) |
2.00 mA … 19.99 mA | 0.01 mA | ±5 % of reading | ||
^{6} Power (active) | P | 0 W … 999 W 1.00 kW … 3.70 kW |
1 W 10 W |
±(5 % of reading + 5 D) ±5 % of reading |
Power Power (active) |
P | 0 W … 999 W | 1 W | ±(5 % of reading + 5 D) |
1.00 kW … 3.70 kW | 10 W | ±5 % of reading | ||
Power (apparent) | S | 0 VA … 999 VA | 1 VA | ±(5 % of reading + 5 D) |
1.00 kVA … 3.70 kVA | 10 VA | ± 5 % of reading | ||
Power (reactive) | Q | ±(0 VAr … 999) VAr | 1 VAr | ±(5 % of reading + 5 D) |
±(1.00 kVAr … 3.70) kVAr | 10 VAr | ±5 % of reading | ||
Power factor | PF | 0.00i … 1.00i | 0.01 | ±(5 % of reading + 5 D) |
0.00c … 1.00c | ||||
Total Harmonic Distortion (voltage) | THDU | 0.0 % … 99.9 % | 0.1 % | ±(5 % of reading + 5 D) |
Total Harmonic Distortion (current) | THDI | 0.00 A … 16.00 A | 0.01 A | ±(3 % of reading + 5 D) |
Cosinus fi | Cos fi | 0.00i … 1.00i | ||
0.00c … 1.00c | 0.01 | ±(5 % of reading + 5 D) | ||
Current | I | 0.00 A … 16.00 A | 0.01 A | ±(3 % of reading + 5 D) |
Voltage | U | 0.0 V … 199.9 V | 0.1 V | ±(3 % of reading + 10 D) |
200 V … 264 V | 1 V | ±3 % of reading | ||
Leak’s & Power ^{7} Power (active) |
P | 0 W … 999 W | 1 W | ±(5 % of reading + 5 D) |
1.00 kW … 3.70 kW | 10 W | ±5 % of reading | ||
^{7} Touch leakage current | Itou | 0.000 mA … 1.999 mA | 1 µA | ±(3 % of reading + 3 D) |
2.00 mA … 19.99 mA | 0.01 mA | ±5 % of reading | ||
Differential leakage current | Idiff | 0.000 mA … 1.999 mA | 1 µA | ±(3 % of reading + 3 D) |
2.00 mA … 19.99 mA | 0.01 mA | ±5 % of reading | ||
Power (apparent) | S | 0 VA … 999 VA | 1 VA | ±(5 % of reading + 5 D) |
1.00 kVA … 3.70 kVA | 10 VA | ± 5 % of reading | ||
Power (reactive) | Q | ~(0 VAr … 999) VAr | 1 VAr | ±(5 % of reading + 5 D) |
~(1.00 kVAr … 3.70) kVAr | 10 VAr | ±5 % of reading | ||
Power factor | PF | 0.00i … 1.00i | 0.01 | ±(5 % of reading + 5 D) |
0.00c … 1.00c | ||||
Total Harmonic Distortion (voltage) | THDU | 0.0 % … 99.9 % | 0.1 % | ±(5 % of reading + 5 D) |
Total Harmonic Distortion (current) | THDI | 0.00 A … 16.00 A | 0.01 A | ±(3 % of reading + 5 D) |
Cosinus fi | Cos fi | 0.00i … 1.00i | 0.01 | ±(5 % of reading + 5 D) |
0.00c … 1.00c | 0.01 A | |||
Current | I | 0.00 A … 16.00 A | 0.1 V | ±(3 % of reading + 5 D) |
Voltage | U | 0.0 V … 199.9 V | 1 V | ±(3 % of reading + 10 D) |
200 V … 264 V | ±3 % of reading | |||
Clamp current ^{8} Clamp current |
Idiff | 0.10 mA … 9.99 mA | 0.01 mA | ±(5 % of reading + 10 D) |
Ip | 10.0 mA … 99.9 mA | 0.1 mA | ±(5 % of reading + 5 D) | |
100 mA … 999 mA | 1 mA | ±(5 % of reading + 5 D) | ||
1.00 A … 9.99 A | 0.01 A | ±(5 % of reading + 5 D) | ||
10.0 A … 24.9 A | 0.1 A | ±(5 % of reading + 5 D) | ||
Insulation Resistance Riso LN-PE, LN-NEP, LN-AP, AP-PE, AP-NEP Riso |
Riso | 0.00 MΩ … 19.99 MΩ | 0.01 MΩ | ±(3 % of reading + 2 D) |
20.0 MΩ … 199.9 MΩ | 0.1 MΩ | ±5 % of reading | ||
^{9} Output voltage | Um | 0 V … 600 V | 1 V | ±(3 % of reading + 2 D) |
Equipment leakage (alternative, direct, differential) ^{10} Equipment leakage current (direct, differential, alternative) |
Ieq | 0.000 mA … 1.999 mA | 1 µA | ±(3 % of reading + 3 D) |
^{10} Ulpe (direct, differential, alternative) | Ulpe | 2.00 mA … 19.99 mA | 0.01 mA | ±5 % of reading |
^{10} Power (direct, differential) | P | 0 V … 299 V | 1 V | ±(2 % of reading + 2 D) |
0 W … 999 W | 1 W | ±(5 % of reading + 5 D) | ||
1.00 kW … 3.70 kW | 10 W | ±5 % of reading | ||
Applied Part leakage (alternative, direct) ^{11} Applied Part leakage current (direct, alternative) |
Iap | 0.000 mA … 1.999 mA | 1 µA | ±(3 % of reading + 3 D) |
2.00 mA … 19.99 mA | 0.01 mA | ±5 % of reading | ||
^{11} Uap (direct, alternative) | Uap | 0 V … 299 V | 1 V | ±(2 % of reading + 2 D) |
^{11} Power (direct) | P | 0 W … 999 W | 1 W | ±(5 % of reading + 5 D) |
1.00 kW … 3.70 kW | 10 W | ±5 % of reading |