HOME / GW Instek

GDM-8342 & GDM-8341

Features :

  • 50,000 counts, VFD display
  • Dual measurement/Dual display
  • Selectable measurement speeds, the maximum: 40 Readings/s for DCV
  • The basic precision of DC voltage: 0.02%
  • Auto/manual Ranging
  • True RMS (AC,AC+DC) measurements
  • 11 different measurement functions
  • Max./Min., REL, MX+B, 1/X, Ref%, Compare, Hold, dB, dBm
  • Standard USB device interface to connect with a computer
  • Temperature measurement function (GDM-8342 only)
  • USB storage for Data collection (GDM-8342 only)
  • Optional GPIB (factory install for GDM-8342)

Category: Benchtop Digital Multimeter

Tags: GDM 8342, GDM 8341, GDM8342, GDM8341, GW Instek Indonesia, GWINSTEK Indonesia, Distributor GW Instek, Jual GW Instek, Jual Digital Multimeters, Benchtop Digital Multimeters, Multimeter Murah

GW INSTEK rolls out the new generation Dual Measurement Multimeter -- the GDM-834x series, which has two models -GDM-8341 and GDM 8342. Its exceptional features include 50,000 counts, VFD dual-display, 0.02% basic DC voltage accuracy and a USB protocol connector to provide users with measurement precision, lucid data observation, and the convenience to connect with the personal computer.
The GDM-834x series not only supports the fundamental measurement items provided by general multimeters, but also equips with capacitance and temperature measurement functions. Furthermore, the GDM-834x series also provides many auxiliary functions to meet the measurement requirements for manufacturing process tests, educational experiments and testing facilities.
With respect to storing and retrieving data, the GDM-834x series has two methods to offer: first, the USB flash drive storage function --- operating alone without connecting with a computer; second, USB interface (virtual COM port) and optional GPIB interface (must be installed in factory) for automatic measurement system users to conveniently save and retrieve data.

Ordering Info :

GDM-8342 with GPIB
50,000 Counts Digit Dual Measurement Multimeter with GPIB interface

GDM-8342
50,000 Counts Digit Dual Measurement Multimeter

GDM-8341
50,000 Counts Digit Dual Measurement Multimeter

GDM-8342 / GDM-8341 Specifications
The specifications apply when the GDM-8342 / GDM-8341 is powered on for at least 30 minutes under +20°C~+30°C

Function Range (*2) Resolution Input Resistance
etc.
Accuracy(*1)
1 Year 23°C±5°C
DC Voltage 500.00 mV 10μV 10MΩ or >10GΩ 0.02 + 4
5.0000 V 100μV 10MΩ or >10GΩ 0.02 + 4
50.000 V 1mV 11.11MΩ 0.02 + 4
500.00 V 10mV 10.1MΩ 0.02 + 4
1000.0 V 100mV 10MΩ 0.02 + 4
Resistance 500.00 Ω 10mΩ 0.83mA 0.10 + 5 (*3)
5.0000 kΩ 100mΩ 0.83mA 0.10 + 3 (*3)
50.000 kΩ 83μA 0.10 + 3
500.00 kΩ 10Ω 8.3μA 0.10 + 3
5.0000 MΩ 100Ω 830nA 0.10 + 3
50.000 MΩ 1kΩ 560nA//10 MΩ 0.30 + 3
DC Current 500.00 μA 10nA 0.06V max. 0.05 + 5
5.0000 mA 100nA 0.6V max. 0.05 + 4
50.000 mA 1μA 0.14V max. 0.05 + 4
500.00 m A 10μA 1.4V max. 0.10 + 4
5.0000 A 100μA 0.5V max. 0.25 + 5
10.000 A 1mA 0.8V max. 0.25 + 5
Continuity 5000.0 Ω 100mΩ 0.83mA 0.10 + 5
Diode Test 5.0000 V 100μV 0.83mA 0.05 + 5
Capacitance 5.000 nF: 0.5~1nF 0.001nF 8.3μA 2.00 + 20
5.000 nF: 1~5nF 0.001nF 8.3μA 2.00 + 10
50.00 nF: 5~10nF 0.01nF 8.3μA 2.00 + 30
50.00 nF: 10~50nF 0.01nF 8.3μA 2.00 + 10
500.0 nF 0.1nF 83μA 2.00 + 4
5.000 μF 1nF 0.56mA 2.00 + 4
50.00 μF 10nF 0.83mA 2.00 + 4
True RMS Volatge 
AC (or AC+DC) (*4/*5/*6)
500.00 mV 10μV 30Hz – 50Hz 1.00 + 40
50Hz – 10kHz 0.50 + 40
10kHz – 30kHz 2.00 + 60
30kHz – 100kHz 3.00 + 120
5.0000 V 100μV 30Hz – 50Hz 1.00 + 20
50Hz – 10kHz 0.35 + 15
10kHz – 30kHz 1.00 + 20
30kHz – 100kHz 3.00 + 50
50.000 V 1mV 30Hz – 50Hz 1.00 + 20
50Hz – 10kHz 0.35 + 15
10kHz – 30kHz 1.00 + 20
30kHz – 100kHz 3.00 + 50
500.00 V 10mV 30Hz – 50Hz ------
50Hz – 10kHz 0.50 + 15
10kHz – 30kHz 1.00 + 20
30kHz – 100kHz 3.00 + 50
750.0 V 100mV 30Hz – 50Hz ------
50Hz – 10kHz 0.50 + 15
10kHz – 30kHz ------
30kHz – 100kHz ------
True RMS Current
AC (or AC+DC) (*4/*5/*6)
500.00 μA 10nA 30Hz – 50Hz 1.50 + 50
50Hz – 2kHz 0.50 + 40
2kHz – 5kHz 1.50 + 50
5kHz – 20kHz 3.00 + 75
5.0000 mA 100nA 30Hz – 50Hz 1.50 + 40
50Hz – 2kHz 0.50 + 20
2kHz – 5kHz 1.50 + 40
5kHz – 20kHz 3.00 + 60
50.000 mA 1μA 30Hz – 50Hz 1.50 + 40
50Hz – 2kHz 0.50 + 20
2kHz – 5kHz 1.50 + 40
5kHz – 20kHz 3.00 + 60
500.00 mA 10μA 30Hz – 50Hz 1.50 + 40
50Hz – 2kHz 0.50 + 20
2kHz – 5kHz 1.50 + 40
5kHz – 20kHz 3.00 + 60
5.0000 A 100μA 30Hz – 50Hz 2.00 + 40
50Hz – 2kHz 0.50 + 30
10.000 A 1mA 30Hz – 50Hz 2.00 + 40
50Hz – 2kHz 0.50 + 30
Frequency 10Hz – 500Hz ------ ------ 0.01 + 5
500Hz – 500kHz 0.01 + 3
500kHz – 1MHz 0.01 + 5
Temperature (*7)
(Thermocouple)
-200 °C ~ +300 °C 0.1 °C J / T / K 2 °C (typical)
Display VFD, Tow Colors Display
Interface USB Device, USB Host (for GDM-8342 only), GPIB (optional for GDM-8342 only)
Power Source AC 100 V / 120 V / 220 V / 240 V ±10%, 50-60Hz
Power Consumption Max. 15VA
Dimensions 265(W) x 107(H) x 302(D) mm
Weight Approx. 2.9kg without option

1. Accuracy = +/- (% of reading + digits)
2. 2% overrange on all ranges, except 10A is 20% overrange
3. REL function is on
4. The AC+DC is in AC coupled
5. The accuracy of AC+DC is equal to AC with 10 more digits added
6. AC characteristics are for sinewave input > 5% of range
7. Specifications do not include probe accuracy

KEEP IN TOUCH